|
As reported by the field, during a distal posterior cerebral artery aneurysm (pca) measuring at approximately 20 mm in size, a microsphere xl 9mmx27cm coil (ssr18092720, l14899) was advanced through a headway duo 156 cm microcatheter (mc) and stretched to the point that the coil detached inside the microcatheter.The physician attempted to use a coil pusher wire to advance the coil through the microcatheter into the aneurysm unsuccessfully.Ultimately, the decision was made to take out the microcatheter.Fortunately, the entire coil was removed from the body.A stryker xt17 was used to reassess the aneurysm and finished coiling with 2 target coils.This is considered a very complicated case due to the location of the aneurysm.Either product was not altered after removal, from the patient¿s body.This did result in the delay of procedure.
|
|
Product complaint # (b)(4).Procode: krd/hcg.The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
|
Product complaint # (b)(4).Complaint conclusion: as reported by the field, during a distal posterior cerebral artery aneurysm (pca) measuring at approximately 20 mm in size, a micrusphere xl 9mmx27cm coil (ssr18092720, l14899) was advanced through a headway duo 156 cm microcatheter (mc) and stretched to the point that the coil detached inside the microcatheter.The physician attempted to use a coil pusher wire to advance the coil through the microcatheter into the aneurysm unsuccessfully.Ultimately, the decision was made to take out the microcatheter.Fortunately, the entire coil was removed from the body.A stryker xt17 was used to reassess the aneurysm and finished coiling with 2 target coils.This is considered a very complicated case due to the location of the aneurysm.Either product was not altered after removal, from the patient¿s body.This did result in the delay of the procedure.No additional information is available.A non-sterile micrusphere xl 9mmx27cm was received contained in the decontamination pouch.Upon receipt of the device, visual inspection was performed, and it was noted that a headway duo (non-j&j) microcatheter was returned.One (1) kink damage was found at 130.1 cm from the proximal end of the microcatheter, and residues of dried blood were observed along the entire length of the microcatheter.The micrusphere delivery unit was not returned for evaluation.The microcatheter was flushed with a lab sample syringe, and resistance was felt.A lab sample guidewire was inserted and advanced through the microcatheter; however, it was found to be obstructed at 37 cm from the proximal end; a dissection was performed at that length, and the proximal end of the micrusphere embolic coil was found stuck and severely stretched.The lab sample guidewire was inserted through the distal end of the microcatheter, and an obstruction was found at 1 cm.Another dissection was performed at that length, and the distal end of the micrusphere embolic coil was found undamaged and covered by a dried coagulum.The issue reported regarding the coil becoming stretched and detached inside the microcatheter was confirmed based on the findings mentioned above.The complaint documented that it was attempted to use a coil pusher wire to advance the coil through the microcatheter into the aneurysm; the location of the proximal end of the coil within the microcatheter suggests that it became stuck and then was pushed with the coil pusher all the way through the distal end of the microcatheter, resulting in the coil stretching, the coil unraveled and pulled free from the delivery system.It is also possible that the coagulum found around the coil occurred due to an adequate flush not being maintained, causing friction and resulting in coil stretching and mechanical detachment/breakage.The location of the aneurysm as reported in the event, and other factors not described, such as vessel characteristics and operator¿s technique, may have contributed to the issue encountered during the procedure.With the limited information available, the root cause remains speculative.A manufacturing record evaluation was performed, and no non-conformances related to the reported complaint condition were identified.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.Devices undergo 100% inspection at different points along the manufacturing process to prevent damages such as coil stretching from leaving the facility.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no capa activity is required at this time.It should be noted that product failure is multifactorial.The instructions for use (ifu) contains the following cautions: if unusual friction is noticed during advancement or retraction of the microcoil system, verify the clear tab is unlocked and pulled out from the re-sheathing tool approximately 1 in (2¿3 cm).If unusual friction is noticed during advancement or retraction of the microcoil system through the introducer, open the rhv main valve, and partially withdraw the distal end of the introducer to expose its tip within the rhv.Tighten the rhv main valve, and flush the y-connector of the rhv with sterile saline and verify that fluid exits the slit in the clear portion of the introducer.After flushing, reinsert the introducer into the infusion catheter hub.If unusual friction is still noticed during advancement or retraction of the microcoil system, verify flush lines are open and properly pressurized.Then slowly withdraw the entire microcoil system and examine for damage.Replace it with a new microcoil system.If friction still exists, withdraw and examine the delivery catheter system.If the microcoil system becomes immobile in the infusion microcatheter, apply a gentle push-pull motion to free it.If unsuccessful, remove both microcatheter and microcoil system together as a unit and replace with new devices.Do not attempt to use the microcoil system as a guidewire if positioning of the microcatheter is lost during microcoil deployment.If repositioning of the microcoil is necessary, carefully observe the motion of the microcoil in respect to the dpu wire while retracting the microcoil under fluoroscopy.If the microcoil movement is not one-to-one with the dpu wire, or if repositioning is difficult, the microcoil may have bome stretched and could possibly break.Gently remove and discard the microcoil system.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|