U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ETRINSA 8 DR-T; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ETRINSA 8 DR-T; PACEMAKER Back to Search Results
Model Number 394931
Medical Device Problem Code Failure to Interrogate (1332)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/10/2023
Type of Reportable Event Malfunction
Event or Problem Description
Device is unable to be interrogated.Device remains implanted.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
The device was not returned for analysis.The analysis is therefore only based on the returned device data as well as the inspection of the quality documents associated with the manufacture of this particular device.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing.The returned device data were inspected.During the inspection the clinical observation could be confirmed.Further analysis revealed an unexpected battery behavior, which led to the clinical observation.Based on the data available for analysis the root cause for that behavior was not determinable and can only be clarified by an analysis of the device itself.However, it cannot be excluded that this occurrence resulted from a defective component, which may compromise the ability of the device to deliver therapy.Biotronik has informed the health care professional about this analysis result, who decides, based on the patients individual circumstances and medical judgment, if the device will be exchanged.Should further relevant information or the device itself become available, this investigation will be updated and you will be informed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETRINSA 8 DR-T
Common Device Name
PACEMAKER
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key16536938
Report Number1028232-2023-01246
Device Sequence Number6042769
Product Code NVZ
UDI-Device Identifier04035479131821
UDI-Public(01)04035479131821(17)160831
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P950037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2016
Device Model Number394931
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2023
Supplement Date Received by Manufacturer03/27/2023
Initial Report FDA Received Date03/14/2023
Supplement Report FDA Received Date03/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/11/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-