| Model Number |
45883 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was reported that the device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
|
| |
|
Event or Problem Description
|
|
A distributor reported an end user experienced an issue with a picc from a bio-stable 5f dl-55cm ir-145 kit,valved with nitinol gw.It was reported the luer lock connectors on a patient's picc line broke.The device was removed and replaced with a new of the same device.The patient did not experience any adverse effects, harm, or require medical intervention because of this incident.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's reported complaint description of picc hub cracked cannot be confirmed, no sample was returned.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.The most likely root cause for the cracked female valve housing is due to an over tightened connection with a mating male luer (likely a needleless connector).A contributing factor to the over-tightened connection may be the mechanics/hand placement during infusion access.It is preferable to grasp the needleless connector (nc) when connecting a syringe/tubing to it rather than grasping the pasv valve luer hub while making a connection with the nc.Grasping the pasv valve hub while connecting a syringe/tubing set to the nc can transmit additional torque to the pasv female luer hub.Inadequate flushing of the device lumen may also be a contributing factor.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint. the review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the dfu that is supplied in bioflo kits contains the following precaution: it is recommended that only luer lock accessories be used with the · bioflo¿ picc with endexo¿ and pasv¿ valve technology.Repeated over-tightening may reduce hub connector life.Do not use hemostats to secure or remove devices with luer lock hub connections.Flushing - recommended procedure 1.Flush the catheter after every use, or at least every seven days when not in use, to maintain patency.Use a 10 ml syringe or larger.2.Flush the catheter with a minimum of 10 ml of sterile normal saline, using a "pulse" or "stop/start" technique.Warning: if using bacteriostatic saline, do not exceed 30 ml in a 24-hour period.3.Disconnect the syringe and attach a sterile end cap to each luer lock hub.Note: this is the recommended flush procedure for this catheter.If using a different procedure than listed above, the use of heparin may be necessary.Follow institutional protocol for catheter flushing.Precaution: incompatible drug delivery within the same lumen may cause precipitation.Ensure that the catheter lumen is flushed following each infusion.Precaution: if resistance is met when flushing, it is recommended that no further attempts be made.Further flushing may result in catheter rupture.Refer to institutional protocol for clearing occluded catheters.Precaution: place a cap on the hub after use to reduce the risk of contamination.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|
| |
|
Search Alerts/Recalls
|