| Model Number |
BEA28-90/I16-30 |
| Medical Device Problem Code |
Lack of Effect (4065)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Clinical Endoleak (4843)
|
| Date of Event |
02/22/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
|
| |
|
Event or Problem Description
|
|
The patient was being treated for an abdominal aortic aneurysm (aaa) with the implant of an afx2 bifurcated stent graft and afx vela suprarenal.Approximately three and a half (3.5) years post-initial procedure, on (b)(6) 2023, aneurysm enlargement with an unknown endoleak were identified.Reintervention was competed on (b)(6) 2023, with the successful reline with the implant of an afx2 bifurcated stent graft and an afx vela suprarenal.Patient was reported to be doing well post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the incident could not be completed.No medical records nor medical imaging relevant to the reported incident was received by endologix.The user facility was unable to provide this information.Due to the absence of medical records and medical imaging device, use, procedure, and/or anatomy relatedness to this incident could not be evaluated.The final patient status remains unknown.The final patient status was not made available to endologix.No additional investigation of this reported incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar incidents.Device iteration is afx2.H6: investigation finding codes - remove code 3233 h6: investigation conclusion codes - remove code 11.
|
| |
|
Search Alerts/Recalls
|