|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was returned to olympus for inspection, and the evaluation findings are as follows: the clip was detached from the distal end and not returned for investigation.There were no device abnormalities observed such as buckling, or damage found in the coil sheath.The hook protruded from the distal end of the coil sheath when the slider was fully pushed.There were no hook deformities observed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 1 year since the subject device was manufactured.Based on the results of the investigation, it is believed that the report of the ¿clip not detaching from the sheath,¿ was due to the device deploying prematurely.It is likely that the user believed the clip to be stuck in the sheath (without reopening) because of the premature deployment.The premature deployment likely occurred because the hook width was near the upper limit.However, the root cause could not be determined.Regarding the report of ¿the clip detaching and falling into the patient¿s abdominal cavity,¿ the root cause could not be determined.However, the incident likely occurred due to the following mechanism: (1) the clip grasped the tissue.(2) the clip became stuck in the product.(3) the entire product was pushed and pulled because the clip became stuck.(4) the clip almost came off from the tissue because the entire clip was pulled.The clip also came off from the main body of the product due to the tensile load added to the clip.(5) the clip that was barely attached to the tissue came off due to some load.(6) the dislodged clip fell out into the abdominal cavity.Furthermore, the incident described above (clip detaching and falling into the patient¿s abdominal cavity) likely caused a procedural delay.However, the root cause could not be determined.For reference, the steps to release the clip have been provided below: 1.When pulling the slider, the operation wire, the hook and the clip arms are linked together, and they are pulled together in the proximal direction.This motion closes the clip arms.2.When the slider is further pulled after closing the clip arms, the protrusions of the clip arms will move to the outside of the clip pipe.Once it happens, the small protrusions of the clip arms will be fix to the edge of the clip pipe.As a result, the clip cannot be opened or closed.(see fig.2) the clip can be opened and closed by operating the slider to a position where the small protrusions of the clip arms do not come out of the clip pipe.3.When the slider is further pulled after the clip does not open or close, the limiter located in the slider breaks with the noise of snapping.(the user recognizes that clipping was completed by hearing the sound.) 4.After the limiter breaks, the joint between the hook and the clip arms will be unhooked by moving the slider forward.As a result, the clip will be released from the product.(see fig.3) the event can be detected/prevented by following the instructions for use (ifu) which state: ¿operation of this instrument is based on the assumption that open surgery is possible as an emergency measure if the clip cannot be detached from the instrument or if any other unexpected circumstance takes place.In this case, refer to chapter 14, ¿emergency treatment¿.¿ ¿do not withdraw the instrument forcibly or change the angulation of the endoscope when the clip is grasping the tissue and is not released.Doing so may tear tissue inside the body cavity, resulting in patient injury, such as perforation, bleeding, or mucous membrane damage.¿ ¿do not angulate the bending section of the endoscope or withdraw the instrument from the endoscope while the clip is grasping the tissue before being released.This could cause patient injury, such as perforation, bleeding, or mucous membrane damage.¿ ¿do not try to forcibly withdraw the instrument from the endoscope if the clip cannot be detached from the instrument.Forcibly withdrawing the instrument could cause patient injury such as perforation, bleeding, or mucous membrane damage.¿ ¿should the slightest irregularity and/or breakage of the parts be suspected, withdraw the instrument from the endoscope immediately according to the steps of ¿withdrawing the instrument from the endoscope¿.Otherwise, perforation, bleeding, or mucous membrane damage may result.¿ this supplemental report includes an update to h3.Also, new information has been added to h4.Olympus will continue to monitor field performance for this device.
|