This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4315.H10: narrative/data was updated.Zimvie did not receive the implant for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item and lot number along with the applicable device history record (dhr), instructions for use (ifu), and risk file (rmf).Dhr review was completed for the subject lot number 2022010326.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2022010326 for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are torque or speed applied during placement/seating exceeds recommended value, material selection of the implant was not adequate to withstand recommended torque values for placement/seating.Therefore, based on the available information, device malfunction could not be verified, and the reported event was non-verifiable with the information provided.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|