| Model Number |
367899 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® k2 edta (k2e) 10.8mg blood collection tube the rubber stopper remained in tube (stopper/shield separation).This event occurred 27 times.The following information was provided by the initial reporter.The customer stated: "there is an ongoing issue with stopper not being fully removed from bd vacutainer tubes, in this case it is for cat 367899 lot 2166071.".
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information has been provided.The following changes have been made below.B.5.Describe event or problem: it was reported when using the bd vacutainer® k2 edta (k2e) 10.8mg blood collection tube the rubber stopper remained in tube (stopper/shield separation).This event occurred 17 times.The following information was provided by the initial reporter.The customer stated: "there is an ongoing issue with stopper not being fully removed from bd vacutainer tubes, in this case it is for cat 367899 lot 2166071.".
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® k2 edta (k2e) 10.8mg blood collection tube the rubber stopper remained in tube (stopper/shield separation).This event occurred 17 times.The following information was provided by the initial reporter.The customer stated: "there is an ongoing issue with stopper not being fully removed from bd vacutainer tubes, in this case it is for cat 367899 lot 2166071.".
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: bd had not received samples or photos for investigation.Therefore, ten (10) retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to closure separation as all samples met specifications.Ninety (90) retains were visually examined and no issues were observed as all samples met specifications.Additionally, twenty (20) retain samples were sent to franklin lakes for r&d to perform further functional testing.R&d retained sample test results were satisfactory.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode of closure separation.Additionally, the ongoing investigation into customer closure separation (decapping) failures has made progress in the last few months.Bd performed a thorough investigation into this issue utilizing root cause analysis tools.Bd identified two potential root causes, and product was manufactured under different experimental conditions to evaluate the potential causes.The product samples were tested to verify key factors in production that may contribute to the separation of the cap from the stopper in automation lines delivering tubes to analyzers.During root cause analysis sessions and testing, bd identified our internal testing method for stopper function required improvement to better reflect the dynamics and forces seen at the decapping step in tube automation lines.This method has been released and continues to be tested in parallel with historic methods to verify effectiveness in replicating the field failures.To date it has been able to better reflect the performance seen at our customer sites and we are moving to fully validate it for use on our design specifications in the future.To date bd has not been able to identify a definitive root cause, but our r&d team will continue to investigate complaints for this defect.A complaint history review was conducted, and no trends were identified.Based on the severity and occurrence rate of this issue no corrective action will be taken at this time.Complaints received for this device and reported conditions will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® k2 edta (k2e) 10.8mg blood collection tube the rubber stopper remained in tube (stopper/shield separation).This event occurred 17 times.The following information was provided by the initial reporter.The customer stated: "there is an ongoing issue with stopper not being fully removed from bd vacutainer tubes, in this case it is for cat 367899 lot 2166071.".
|
| |
|