• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE; PISTON SYRINGE Back to Search Results
Catalog Number 305271
Medical Device Problem Codes Leak/Splash (1354); Retraction Problem (1536); Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/21/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.No root cause can be determined as no samples were received.The lot number is unknown, therefore device history record review (dhr) or quality notification review (qn) could not be performed.Examination of the product involved may provide clarification as to the cause for the reported failure.We appreciate you taking the time to bring this observation to our attention.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that the bd integra¿ syringe with detachable needle leaked fluid down the patient's arm and was removed.However, when the button was pressed to retract the needle, part of the spring and plastic end "flew off" to the other side of the room.The following information was provided by the initial reporter: "administering im droperidol to patient using auto retractable syringe.While using, fluid leaked out of syringe and down pts arm, unknown how much was actually administered.Removed from pt before retracting due to obvious issue with item.On pressing button to retract needle, part of the item (spring and plastic end) flew off landing on the other side of the room.Managed to visualise needle inside syringe to confirm it had not also been launched across room.As per customer: it is likely that the needle was not checked and secured to the syringe when this incident occurred and we are currently putting it down to human error.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16542567
Report Number1213809-2023-00200
Device Sequence Number6490559
Product Code MEG
UDI-Device Identifier00382903052714
UDI-Public00382903052714
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K011103
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number305271
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2023
Initial Report FDA Received Date03/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-