| Model Number |
M00558630 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0402 captures the reportable event of balloon burst.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre pro wireguided dilatation balloon was used in the biliary tract during endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2023.During the procedure, the balloon burst at 9atm.The customer stated that no pieces of the balloon detached inside the patient.The procedure was completed with another cre pro wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0402 captures the reportable event of balloon burst.Block h10: investigation results: the returned cre pro wireguided dilatation balloon was analyzed, and a visual examination found that the balloon and catheter of the device had no damages.Functional analysis was performed, and the balloon was inflated without a problem.However, the balloon would not hold pressure due to pinhole in the distal section.Microscopic inspection found a pinhole located approximately at 15 mm from the tip of the balloon.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon burst was not confirmed.The pinhole problem found could have been interpreted by the customer as the reported event of balloon burst.The pinhole is likely to have occurred due to factors encountered during the procedure: excess pressure, interaction with other devices, or patient anatomy.It is possible that factors encountered during the procedure, such as the interaction with any surface, could create friction on the balloon and cause a balloon pinhole.Therefore, the most probable root cause is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu)/product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre pro wireguided dilatation balloon was used in the biliary tract during endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2023.During the procedure, the balloon burst at 9atm.The customer stated that no pieces of the balloon detached inside the patient.The procedure was completed with another cre pro wireguided dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
| |
|