MEDTRONIC MEXICO BIO-MEDICUS NG ARTERIAL CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
| Model Number |
96570-121 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information that prior to use of a bio-medicus nextgen femoral arterial or jugularvenous cannula, it was reported that when preparing for insertion of the device, a crack was discovered on the connector of the cannula. there had been no external impact on the cannula and the crack was detected before insertion into the patient. the device was replaced to complete the procedure. there was no patient involvement so no adverse effect occurred.Medtronic received additional information that the cannula was not heated or cooled prior to use.There was no patient blood loss or transfusion required as the crack was detected before insertion into the patient.There was no damage to the packaging (outer box, inner packaging or sterile barrier).No other devices in the same box/shipping container were damaged.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual inspection showed evidence of damage/cracks in the connector.The hemostasis cap was not returned.The reason for return was confirmed.Conclusion: after investigation the complaint is confirmed for a cracked cannula body connector.Returned product analysis confirmed the damage to the connector and the hemostasis cap was not returned with the cannula.It is possible, based on the photo provided by the customer, that the hemostasis cap was not used when the introducer was placed into the cannula body prior to use and therefore, this complaint would most likely be the result of a use-related issue.This damage to the cannula connector can occur when the introducer is inserted into the cannula body without the use of the hemostasis cap.The introducer handle can cause stretching, crazing, or cracking when it is forced into the connector.The issue is more likely to occur with the arterial cannulae models with a clear hemostasis cap that is attached to the introducer on the provided protective sheath, however, the issue can also occur in the venous cannulae models with the blue hemostasis cap.Review of the device history record was not completed as there is no evidence to sug gest manufacturing issues with the complaint device.A manufacturing investigation was completed for this failure mode, in this investigation the manufacturing process was reviewed, and the handle of the affected part is minimum, just pick and place into the packaging, no additional assembly.Review of in-process pictures showed that the pieces produced are in good condition.Product event notification and awareness was given to the production personnel about the implications and consequences that reported defect has on the field.Complaints received for similar model numbers were reviewed and showed no trends warranting escalation related to this occurrence.There were no adverse patient effects because of this incidence.Medtronic will continue to monitor for future occurrences and trends.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|