H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: received one bard magnum biopsy needle.Upon visual evaluation the device appeared to be clean.The needle and sample notch were noted to be bent.No other visual anomalies were noted to the device.Functional testing was not performed, due to the nature of the complaint.Therefore, the investigation for the reported needle bent is confirmed as the needle bent was noted in the physical evaluation.However, the reported difficult to remove remains inconclusive as the condition of use could not be replicated.A definitive root cause for the alleged needle bent and difficult to remove could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 08/2024), g3.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|