| Model Number |
MI1000 MED-EL CONCERT |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user experienced a worsening of the hearing performance with the device.Re-implantation is considered.
|
| |
|
Event or Problem Description
|
|
The user experienced a worsening of the hearing performance with the device.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is plausible.However, to determine an exact root cause a device investigation of the explanted device is necessary.The concerned device was explanted but has not been received for investigation yet.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user experienced a worsening of the hearing performance and fluctuation of loudness with the device.The user has been re-implanted.
|
| |
|
Search Alerts/Recalls
|