Model Number UHI-4
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2023
Event Type  malfunction  
Event Description
The customer reported that his nurse inspected his olympus high flow insufflation unit during procedure preparation and found it was experiencing a power issue.According to the initial reporter, after pressing the power button, the unit would turn on, the immediately turn itself off.Reportedly, a back-up unit was used to perform intended procedure.
 
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user¿s report was confirmed.A faulty printed circuit board was discovered.This board could no longer control it¿s functions and was causing the unit¿s alarm to sound.Additionally, the actual flow rate value could not reach the set flow rate due to the manifold unit being defective.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Correction to h4.Information was inadvertently not included on the initial medwatch.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 6 years since the subject device was manufactured.Based on the results of the investigation, and since the device malfunction of the power suddenly turning off was not confirmed during evaluation, the definitive root cause of the reported issue could not be determined.However, a probable cause is likely due to a faulty printed circuit board.Based on the results of the investigation, it is likely any function could not be controlled, and the alarm sound wouldn¿t stop occurred due to a fault printed circuit board.However, a definitive root cause could not be determined.Based on the results of the investigation, it is likely the actual flow rate value not reaching the setting flow rate value occurred due to a faulty manifold unit.However, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16545530
MDR Text Key311600845
Report Number3002808148-2023-02512
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeVM
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-