| Model Number |
1506-00-007 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Codes |
Adhesion(s) (1695); Foreign Body Reaction (1868); Unspecified Infection (1930); Pain (1994); Synovitis (2094); Joint Laxity (4526); Unspecified Musculoskeletal problem (4535); Unspecified Tissue Injury (4559)
|
| Date of Event |
12/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Component code: appropriate term/code not available (g07002) used to capture no findings available.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient underwent a right attune primary knee arthroplasty on (b)(6) 2015.Patella was resurfaced and competitor cement was utilized.Patient was later revised due to pain on (b)(6) 2022.Prior imaging was consistent with aseptic loosening of the tibial component.Intraoperative findings included grayish staining of the tissue consistent with some mechanical motion of the components.After the knee was flexed to 90 degrees, the tibia was noted to be entirely subluxated anteriorly.The pcl was ¿certainly stretched¿ at this point indicating a soft tissue injury and instability.There was scar and bony growth around both the posteromedial and posterolateral portions of the tibia.The tibial component was popped completely free from the cement with minimal pressure from an osteotome.Scar tissue was removed circumferentially around the patella.The patella had minimal wear and was retained.Doi: (b)(6) 2015, dor: (b)(6) 2022, (right knee).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: a2 age, a3, b5, b7, e1 (facility name), h4, h6 health effect - clinical code.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: a1, d10 concomitant.
|
| |
|
Event or Problem Description
|
|
Medical records received.On (b)(6) 2015, the patient had a right knee arthroplasty to address arthritis.Depuy components along with competitor cement was implanted during the procedure.On (b)(6) 2022, the patient had a revision right total knee to address failed right total knee.The indications for surgery included persistent pain, suspicion for infection with inflammatory markers and aspiration; however, imaging consistent with aseptic loosening of the tibial component.During the procedure, the surgeon observed some grayish staining of the tissue consistent with some mechanical motion of the components.Tibial component popped free from cement consistent with loosening seen consistently with first generation attune tibias.Depuy components were implanted during this procedure including depuy cement.
|
| |
|