| Model Number |
A2025-120 |
| Medical Device Problem Codes |
Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.Device code 2017 - incorrect prepna.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation the 2.5/120 mm armada 14 percutaneous transluminal angioplasty (pta) catheter was inflated and the device would not fill up prior being used on the patient.Another armada was used to complete the procedure.There was no device use or patient involvement, and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.It was reported that the balloon was inflated during preparation and before entering the patient anatomy.It should be noted that the percutaneous transluminal angioplasty (pta) catheter, armada 14, global, instructions for use (ifu) states the steps to air aspirate the device and slowly release device pressure to neutral.In this case, it is unknown if the ifu violation caused or contributed to the reported complaint since the device was not returned for analysis.The investigation was unable to determine a conclusive cause for the reported complaint.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.D10, h3 - it was initially reported that the device would be returned for analysis.Subsequent information revealed that the device was discarded and is not available for evaluation.
|
| |
|
Search Alerts/Recalls
|