• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURNEY II UK TIB INSRT MED SZ 5-6 9MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURNEY II UK TIB INSRT MED SZ 5-6 9MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 74026169
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).")).This complaint was opened by smith+nephew to document a patient complication identified through the review of clinical evidence from post market clinical data collection activities that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that, after a left medial unicompartmental knee replacement had been performed on an unspecified date to address ahlbäck disease symptoms, the patient required a revision surgery on an unknown date due to an unspecified infection where the tibial insert was exchanged.According to the data set provided, this event occurred seven (7) days following the index implantation.The outcome of this patient is unknown.This incident was noticed in a retrospective post-market clinical follow up activity (pmcf) where anonymized data summarizing outcomes were gathered retrospectively; therefore, additional information is not known.
 
Additional Manufacturer Narrative
D4: catalog number and unique identifier (udi) # ,d11: concomitant medical products and therapy dates and g5: 510(k) #.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY II UK TIB INSRT MED SZ 5-6 9MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16546634
Report Number1020279-2023-00537
Device Sequence Number6389164
Product Code HSX
UDI-Device Identifier03596010582218
UDI-Public03596010582218
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K061011
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number74026169
Device Catalogue Number71422262
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2023
Supplement Date Received by Manufacturer03/08/2023
Initial Report FDA Received Date03/15/2023
Supplement Report FDA Received Date03/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
JOURNEY II UK MED TIB BASE SZ 6 LT.; JOURNEY II UK OX FEM SZ 6 LM/RL.; JRNY UNI TIBIAL BASE LM/RL SZ 6.; JUNI COCR FB FEM SZ 6 LM RL.
Outcome Attributed to Adverse Event Other;
Patient Age69 YR
Patient SexMale
-
-