• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 DEPUY1 GENTAMICIN CEMENT 40G; BONE CEMENT : BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 DEPUY1 GENTAMICIN CEMENT 40G; BONE CEMENT : BONE CEMENT Back to Search Results
Model Number 5450-31-500
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/28/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision completed due to aseptic loosening at the cement to implant interface.The femur was not loose but after striking the tibia four or five times it came out.Surgeon decided to remove the femur as well and revise both sides since patient had soft bone.No surgical delay.Doi: (b)(6) 2022.Dor: (b)(6) 2023.Affected side : right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPUY1 GENTAMICIN CEMENT 40G
Common Device Name
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16548024
Report Number1818910-2023-05909
Device Sequence Number20446295
Product Code MBB
UDI-Device Identifier10603295174264
UDI-Public10603295174264
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K081163
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5450-31-500
Device Catalogue Number545031500
Device Lot Number9647102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/28/2023
Supplement Date Received by Manufacturer04/05/2023
Initial Report FDA Received Date03/15/2023
Supplement Report FDA Received Date04/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE PS FEM RT SZ 6 CEM; ATTUNE PS RP INSRT SZ 6 12MM; ATTUNE RP TIB BASE SZ 6 CEM; DEPUY1 GENTAMICIN CEMENT 40G; DEPUY1 GENTAMICIN CEMENT 40G
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-