| Model Number |
10-80-00Z |
| Medical Device Problem Code |
Pumping Stopped (1503)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There was no patient involvement.Livanova deutschland manufactures the s5 roller pump pro.The incident occurred in san antonio, texas.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.In order to fix it, motor control board and front panel of the s5 roller pump pro were replaced.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova deutschland received a report that a s5 roller pump pro gave two motor control failure error messages at setup.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the involved pump is installed since 2021 and according to the review of the complaints database, no further event has been reported for this unit.Based on the information available, it cannot be excluded that the root cause could be traced back to an electrical/electronic deviation involving the replaced parts.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|