• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE EF 12 MM THICKNESS; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE EF 12 MM THICKNESS; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the instrument was returned missing bearings on a worn instrument claim form.There was no known patient involvement.
 
Additional Manufacturer Narrative
(b)(4).The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-00704 0001822565-2023-00705 0001822565-2023-00706 0001822565-2023-00707 0001822565-2023-00708 0001822565-2023-00709 0001822565-2023-00710 0001822565-2023-00711 0001822565-2023-00712 0001822565-2023-00713 0001822565-2023-00714 0001822565-2023-00715 0001822565-2023-00716 0001822565-2023-00720 0001822565-2023-00722 0001822565-2023-00723 0001822565-2023-00724 0001822565-2023-00725.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.  visual examination of the returned product found them to exhibit signs of repeated use and have one or both of the detent balls(ball bearings) disassembled/missing the components.The complaint is confirmed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Supplier dhr¿s were reviewed for deviations and/or anomalies with no deviations or non-conformance's found reference attached supplier review for details.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.Reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The complaint is confirmed.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE EF 12 MM THICKNESS
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16552694
Report Number0001822565-2023-00719
Device Sequence Number5088578
Product Code JWH
UDI-Device Identifier00889024245860
UDI-Public(01)00889024245860(10)64254201
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42527900502
Device Lot Number64254201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2023
Supplement Date Received by Manufacturer04/11/2023
Initial Report FDA Received Date03/16/2023
Supplement Report FDA Received Date04/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/04/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-