| Model Number |
N/A |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Codes |
Bacterial Infection (1735); Failure of Implant (1924); Pain (1994); Osteolysis (2377); Swelling/ Edema (4577); Joint effusion (4849)
|
| Date of Event |
02/22/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a left knee arthroplasty revision to address pain, stiffness and infection approximately seven (7) months post-operatively.Attempts have been made, however, no additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: persona trabecular metal cruciate retaining standard porous femoral component, catalog #: 42502806601, lot #: 65353953.Persona medial congruent articular surface left 10mm, catalog #: 42512100810, lot #: 65053538.Nexgen trabecular metal standard primary patella, catalog #: 00587806538, lot #: 64975784.Report source - foreign: event occurred in australia.The complainant has indicated that the product will not be returned to zimmer biomet for investigation due to hospital policy.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this patient; please see all reports associated with this event: 0001822565-2023-00693, 3007963827-2023-00050 and 0001822565-2023-00694.Investigation incomplete.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Review of the sterile certifications found the devices to be conforming with no applicable deviations.The devices were verified to have gone through acceptable sterilization following iso/aami/astm and eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet such as hospital/surgical environment, provider-related risk factors and/or patient comorbidities.As there were no indications of a product or process issue identified affecting implant safety or effectiveness, the implanted products are not identified to have contributed to the reported infection.
|
| |
|
Search Alerts/Recalls
|