STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_13.2
Medical Device Problem Codes Misfocusing (1401); Inadequacy of Device Shape and/or Size (1583); Optical Problem (3001)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Disturbances (2140); Halo (2227); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Weight: unk.Ethnicity: unk.Race: unk.Health impact- clinical code: 4581 - significant reduction of irido-corneal angles work order search: no similar complaint was reported for units within the same lot.Claim # (b)(4).
 
Event or Problem Description
The reporter indicated the surgeon implanted a 13.2mm vticm5_13.2 implantable collamer lens, -05.50/+2.5/095 (sphere/cylinder/axis), into the patients right eye (od) on (b)(6) 2023.The lens was removed on (b)(6) 2023 due to excessive vaulting, significant reduction of irido-corneal angles, glare/haloes and blurred vision.The lens was replaced with a shorter lens and the problem was resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
MDR Report Key16555168
Report Number2023826-2023-00861
Device Sequence Number16852698
Product Code QCB
UDI-Device Identifier00840311311934
UDI-Public00840311311934
Combination Product (Y/N)N
Initial Reporter CountryLG
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model NumberVTICM5_13.2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2023
Initial Report FDA Received Date03/16/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL SFC-45 - LOT # 1592708.; FOAM TIP PLUNGER MODEL FTP - LOT # 1585298.; INJECTOR MODEL MSI-PF - LOT # UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age25 YR
Patient SexFemale
-
-