| Model Number |
10101 |
| Medical Device Problem Code |
Environmental Particulates (2930)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported an event of a ¿white smoke¿ or mist emitting from the sterrad® 100s sterilizer.The customer powered off the unit and discontinued use.There was no report of any injuries or human reactions.This event is being reported as a malfunction report subsequent to a previous serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter phone number: (b)(6).The customer has not provided facility access for asp field service engineer to assess the unit; therefore, details of the required service and resolution are unknown.The device history record (dhr) was reviewed and no issues relating to the failure mode were noted.The involved unit met manufacturer specifications at the time of release.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by advanced sterilization products, or its employees that the report constitutes an admission that the product, advanced sterilization products, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Asp complaint ref #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A field service engineer was dispatched to the customer site.The adapter converter, catalytic converter, oil mist filter, vacuum pump oil, and the vacuum pump were replaced to resolve the smoke/mist issue.Unit meets specifications and was returned to service.H3: asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the smoke/mist issue, and system risk analysis (sra).¿ trending analysis of the smoke/mist issue for the sterrad® 100s unit was reviewed for the prior six months from the event date, and no significant trend was observed.¿ review of risk documentation shows the risk for exposure to toxic or corrosive material to be "low." no parts were available for further analysis.The assignable cause of the smoke/mist is potentially due to the adapter converter, catalytic converter, oil mist filter, vacuum pump oil, and the vacuum pump.The asp field service engineer replaced the parts and confirmed the sterrad® 100s met functional specifications after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
|
| |
|
Search Alerts/Recalls
|