Model Number N/A
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Occupation: director of logistics.Pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported a cook neff percutaneous access set was contaminated.A particle was found inside the packaging of the device by the distributor.No patient contact was made.Therefore, no adverse effects occurred.
 
Event or Problem Description
No additional information regarding patient and/or event details has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged or unavailable.Investigation ¿ evaluation ifs miami (colombia) contacted cook on (b)(6) 2023 concerning a neff percutaneous access set (rpn: npas-104-rh-nt, lot# 15037452).The distributor found foreign matter in the package on (b)(6) 2023.No patient contact was made.Reviews of the documentation, including the complaint history, device history record and quality control procedures, as well as a visual inspection of the returned device, were conducted during the investigation.One unopened neff percutaneous access set was returned for evaluation.Upon a visual inspection of the device, a foreign particle was found near the distal tip of the dchn needle.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that process steps were identified to ensure this nonconformance does not leave the house.A review of the device history record (dhr) for lot 15037452 found no nonconformances that could have contributed to the reported failure mode.It should be noted that there were no other complaints associated with the final product lot number.Cook was unable to review product labeling, as this device is not supplied with an instruction for use (ifu) pamphlet.Evidence gathered from a review of the dmr, dhr and device failure analysis suggests that the device was manufactured out of specification.There is no evidence of nonconforming material in house or in the field.Based on the information provided, examination of the returned product, and the results of our investigation, a definitive cause for the failure could be attributed to a quality control deficiency.The appropriate personnel have been notified.Per the risk assessment no further action is required.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEFF PERCUTANEOUS ACCESS SET
Common Device Name
KGZ ACCESSORIES, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key16558821
Report Number1820334-2023-00266
Device Sequence Number17498560
Product Code KGZ
UDI-Device Identifier00827002105461
UDI-Public(01)00827002105461(17)251025(10)15037452
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date10/25/2025
Device Model NumberN/A
Device Catalogue NumberNPAS-104-RH-NT
Device Lot Number15037452
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/14/2023
Supplement Date Received by Manufacturer06/07/2023
Initial Report FDA Received Date03/16/2023
Supplement Report FDA Received Date06/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-