| Model Number |
29MJ-501 |
| Medical Device Problem Code |
Incomplete Coaptation (2507)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2023, a 29mm sjm masters series mechanical heart valve was selected for an implant.During the procedure, but prior to implant, the physician felt a sense of incongruity, therefore avoided using the device.It was observed opening and closing the leaflet was not as smooth as usual.Device was replace with a (epic 31mm) was used and the procedure was completed.There were no adverse patient effects.The patient is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of difficulty in opening and closing the valve leaflets was reported.Information from the field indicated that the reported event was noticed during the device preparation before the valve implant procedure.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to the labeling design or manufacturing of the device.
|
| |
|