• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIG45CTAV
Medical Device Problem Codes Misfire (2532); Failure to Fire (2610); Component or Accessory Incompatibility (2897); Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2023
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a video-assisted thoracoscopic lobectomy, during vascular resection, the cartridge could not be loaded properly.An approach was made, but the device could not be fired.A new cartridge was used, but the cartridge did not come off either.The surgeon replaced both the handle and reload to resolve the issue.There was no patient injury.Medtronic's initial evaluation of the incident is that the reload was partially fired and the interlock was engaged.The adapter was damaged and bent.
 
Additional Manufacturer Narrative
Concomitant products: egiaushort egiaushort endogia ultra univ sht stap lot #:unknown.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the reload was partially fired and the interlock was engaged.The adapter was damaged and bent.It was reported that the device was difficult to load and unload.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: ensure that the black return knob on the instrument is pulled back completely and the articulation lever is neutral (0° relative) to the instrument.Cycle the instrument to confirm proper loading of the reload.To do so, squeeze the handle once to close the jaws of the reload.Push the black loop handle all of the way forward or pull back on the black return knob and confirm that the reload jaws open fully.Failure to completely fire the reload will result in an incomplete cut and/or incomplete staple formation, which may result in poor hemostasis and/or leakage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16564681
Report Number1219930-2023-01026
Device Sequence Number5535676
Product Code GDW
UDI-Device Identifier10884521543706
UDI-Public10884521543706
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160176
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSIG45CTAV
Device Catalogue NumberSIG45CTAV
Device Lot NumberN2J0361Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2023
Initial Report FDA Received Date03/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-