|
Product complaint #(b)(4).Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that on an unknown date during a procedure, it appeared that the fenestrated lag screw missed the tfna nail when using the 130 degree aiming arm.The lag screw ended up missing the nail anterior causing the surgeon to free hand the lag screw in the correct place.Also, the tfna helical blade and screw inserter broke while turning the screw.The threaded tip snapped off when inserting.This complaint involves three (3) devices.This report is for one (1) 10/130 deg ti cann tfna 360/right - sile this is report 1 of 3 for complaint (b)(4).
|