| Model Number |
CA 15-3 G2 |
| Medical Device Problem Codes |
Mechanical Problem (1384); High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A sample from the patient was submitted for investigation.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable high results for multiple assays for one patient from cobas e 801 analytical unit serial number (b)(4).The results were reported outside of the laboratory and the doctor believed the results did not match the medical imaging results and questioned if there were interfering substances.Refer to the attachment to the medwatch for all patient data.Refer to the medwatches with the following a1 patient identifiers for all involved assays: (b)(6) for the ca 15-3 assay.(b)(6) for the cea assay.(b)(6) for the cyfra 21-1 assay.(b)(6) for the ca19-9 assay.(b)(6) for the ca 125 assay.(b)(6) for the afp assay.
|
| |
|
Additional Manufacturer Narrative
|
|
The patient sample was investigated and no cause for the result discrepancies could be found.The customer¿s results were confirmed.No product issue was found and the reagent performed within specifications.A general reagent-related issue could be excluded.
|
| |
|
Additional Manufacturer Narrative
|
|
Medwatch fields b3 and b6 were updated.
|
| |
|