| Catalog Number |
07027559190 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with elecsys insulin on a cobas e 801 module (serial number (b)(4)).No questionable results were reported outside of the laboratory.The sample initially resulted in an insulin value of 37.2 uu/ml and it repeated as 19.2 uu/ml.
|
| |
|
Additional Manufacturer Narrative
|
|
The patient sample was repeated because the initial value was high and the patient had no result history.An additional repeat of the sample was performed and this value was similar to the repeat value of 19.2 uu/ml.Calibration and control recoveries were within expectations.The investigation could not identify a product problem.The cause of the event could not be determined.Medwatch field a3 has been updated.
|
| |
|