MAUDE Adverse Event Report: SMITH & NEPHEW, INC. EVOS 3.5MM PM-P TIBIA T-PL 13H R 170MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. EVOS 3.5MM PM-P TIBIA T-PL 13H R 170MM; PLATE, FIXATION, BONE Back to Search Results
Model Number 72463413
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code Perforation (2001)
Date of Event 02/20/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference:(b)(4).
 
Event or Problem Description
It was reported that, during an internal fixation surgery, when a 3.5 cortical screw was placed on power in the top row of the evos 3.5mm pm-p tibia t-pl 13h r 170mm, the screw head drove straight through the plate and went into the patient's bone.The star tab was destroyed and the plate was deemed unusable due to loss of fixation.It is unknown if this incident caused a surgical delay.The device was removed and a direct lateral 13 hole right was used instead.Current health status of patient is unknown.
 
Additional Manufacturer Narrative
The device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, as of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.It is unknown if the instructions for use for the evos plating system was adhered to.This document does warn; ¿it is the surgeon¿s responsibility to be familiar with the appropriate surgical techniques prior to use of this device.¿ the impact to the patient beyond that which has already been reported in the complaint cannot be confirmed nor concluded.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for ifu for evos¿ plating system revealed in the adverse effects section that bending, cracking or fracture of implant components can occur.Besides, the warnings section revealed that it is the surgeon¿s responsibility to be familiar with the appropriate surgical techniques prior to use of this device.Also, the important information section revealed that only components made from the same material shall be used together; dissimilar metals shall not be mixed at any time and components from different manufacturers shall only be mixed when advised by the manufacturer.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include surgical technique used or user/procedural variance.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS 3.5MM PM-P TIBIA T-PL 13H R 170MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16569347
Report Number1020279-2023-00579
Device Sequence Number12296704
Product Code HRS
UDI-Device Identifier00885556640623
UDI-Public00885556640623
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K170457
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number72463413
Device Catalogue Number72463413
Device Lot Number20MM02298
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2023
Supplement Date Received by Manufacturer04/05/2023
Initial Report FDA Received Date03/18/2023
Supplement Report FDA Received Date04/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-