Model Number ACCK5300
Medical Device Problem Codes Insufficient Device Problem Information (3190); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 06/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced pain at the anchor site.The anchor was replaced.The anchor was received damaged with one of the eyelets cut off and missing.The physician didn¿t mention removing it or leaving it in situ during the procedure, however the exact location of the eyelet is unknown.There have been no reports of further complications regarding this event.
 
Additional Manufacturer Narrative
The device was returned and the device evaluation is in progress.The manufacturing records were reviewed and no relevant nonconformities were found.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Additional Manufacturer Narrative
Updated b5 + h6 to capture correction that anchor was removed, not replaced.The device was returned and analyzed.The investigation found that one of the anchor¿s silicone loops was torn.Most likely this occurred during the explant procedure, but this cannot be confirmed.It was reported that the patient¿s anchor site pain has since resolved.The source of the pain could not be determined.The investigation found no functional issue with the return lead anchor.  nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Event or Problem Description
Correction: the anchor was removed, not replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENZA
Common Device Name
NEVRO SENZA
Manufacturer (Section D)
NEVRO CORP.
1800 bridge parkway
redwood city CA 94065
Manufacturer Contact
zulfiqar ali
1800 bridge parkway
redwood city, CA 94065
6505317627
MDR Report Key16569626
Report Number3008514029-2023-00143
Device Sequence Number5895893
Product Code LGW
UDI-Device Identifier00813426020251
UDI-Public00813426020251
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P130022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/09/2022
Device Model NumberACCK5300
Device Catalogue NumberACCK5300
Device Lot Number5004506
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/21/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2023
Supplement Date Received by Manufacturer03/29/2023
Initial Report FDA Received Date03/18/2023
Supplement Report FDA Received Date04/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-