| Model Number |
ACCK5300 |
| Medical Device Problem Codes |
Insufficient Device Problem Information (3190); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Implant Pain (4561)
|
| Date of Event |
06/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient experienced pain at the anchor site.The anchor was replaced.The anchor was received damaged with one of the eyelets cut off and missing.The physician didn¿t mention removing it or leaving it in situ during the procedure, however the exact location of the eyelet is unknown.There have been no reports of further complications regarding this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and the device evaluation is in progress.The manufacturing records were reviewed and no relevant nonconformities were found.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated b5 + h6 to capture correction that anchor was removed, not replaced.The device was returned and analyzed.The investigation found that one of the anchor¿s silicone loops was torn.Most likely this occurred during the explant procedure, but this cannot be confirmed.It was reported that the patient¿s anchor site pain has since resolved.The source of the pain could not be determined.The investigation found no functional issue with the return lead anchor. nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
|
| |
|
Event or Problem Description
|
|
Correction: the anchor was removed, not replaced.
|
| |
|