| Model Number |
AUTOMATE 2550 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
02/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The customer reported that the sorting drive of their automate 2550 is aliquoting the wrong tube and storing the correct tube which resulted in the release of false troponin patient results.The customer reported that there was a change in patient treatment; however, no further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer did not provide information relating to the change in patient treatment.Beckman coulter field service engineer (fse) evaluated the automate 2550.The log files from the instrument were reviewed.The fse and application support manager determined that the files indicate that the customer may have had two tubes from two different patients.The log files did not indicate an issue with the instrument.There is no evidence of a instrument malfunction.The customer was satisfied with the support and no further issues were reported.Initial reporter telephone number is (b)(6).The beckman coulter internal identifier is case (b)(4).
|
| |
|