| Model Number |
ONB5SHF |
| Medical Device Problem Codes |
Break (1069); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during laparoscopic hysterectomy, when inserting the trocar, the inner cylinder could not be removed after the trocar ruptured.A new trocar was used to resolve the issue in order to complete the case.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the obturator was received inserted into the cannula.Prolonged insertion can cause the seals to become stretched.The obturator had to be forcibly removed.The tip of the cannula was gouged and pushed in making removal of the obturator difficult.The circular seal and duckbill seal were stretched out.The stopcock appeared intact.Functional testing found the device failed an air leak test.The product was shipped back with the obturator inserted in the cannula and therefore the length of time it was inserted during shipment may have also resulted in the stretched seals.It was reported that the user experienced difficulty in the insertion or removal of the device to or from the port, and the obturator was damaged.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur as a result of rough handling.Also, the cut in the circular seal may occur when contact is made with a sharp surgical device during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: use special care when introducing or removing sharp-edged or sharp-angled endoscopic instruments to minimize the potential of inadve rtent damage to the seal.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|