• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMPLANT DIRECT SYBRON MANUFACTURING LLC Legacy¿3 Implant; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMPLANT DIRECT SYBRON MANUFACTURING LLC Legacy¿3 Implant; DENTAL IMPLANT Back to Search Results
Catalog Number 865210
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Failure of Implant (1924); Osteopenia/ Osteoporosis (2651)
Date of Event 02/20/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The implant failure modes, failure/ loss of osseo-integration and lack of primary stability are not product related and rather are attributed to patient contraindications, conditions, or clinician error in surgical protocol.No product investigation or corrective actions required.Complaints will continue to be trended.
 
Event or Problem Description
Per complaint (b)(4), after clinical procedure, patient experienced loss of implant to osseointegrate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Legacy¿3 Implant
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
IMPLANT DIRECT SYBRON MANUFACTURING LLC
3050 east hillcrest drive
thousand oaks CA 91362
Manufacturer Contact
sergio alvarado
3050 east hillcrest drive
thousand oaks, CA 91362
8057785975
MDR Report Key16571816
Report Number3001617766-2023-01184
Device Sequence Number5015256
Product Code DZE
UDI-Device Identifier10841307102632
UDI-Public10841307102632
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K090234
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Dentist
Remedial Action Inspection
Type of Report Initial
Report Date (Section B) 03/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number865210
Device Lot Number172952
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/23/2023
Initial Report FDA Received Date03/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
Patient SexFemale
Patient Weight91 KG
-
-