MAUDE Adverse Event Report: STRYKER-ENDOSCOPY LAKELAND 5300; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER-ENDOSCOPY LAKELAND 5300; ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R Back to Search Results
Model Number LF4418
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 02/01/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It is unknown if the device failed to seal, reported an error message on the generator, or failed to cut.The surgeon stated the patient had more bleeding and sutures were required.Extended procedure time was to replace device and suture.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
The device was not returned to stryker sustainability solutions for evaluation.Rep stated all the devices discarded by staff.As the device was not returned for evaluation, inspection was unable to be performed.A review of the dhr could not be performed as the lot number and/or serial number was not reported.The reported event could be attributed to: incompatible software revision on generator.Ancillary equipment failure.Not fully seating device into console during initial connection.Grasping on too much tissue or inappropriate tissue types or on staples or tissue build up (eschar).Device not properly cleaned.Trigger button (activation button) not engaged throughout the entire seal cycle.Device activated in contact with pooling fluid.Device used on vessels thicker than 7 mm.The instructions for use (ifu) state: use caution during surgical cases in which patients exhibit certain types of vascular pathology (atherosclerosis, aneurysmal vessels, etc.).For best results, apply the seal to unaffected vasculature.If the generator provides multiple power settings, use the lowest power needed to achieve the intended effect.Do not use this instrument on vessels in excess of 7 mm in diameter.If the instrument shaft is visibly bent, discard and replace the instrument.Do not place the vessel and/or tissue in the jaw hinge.Place the vessel and/or tissue in the center of the jaws.Eliminate tension on the tissue while sealing and cutting to ensure proper function.Use caution when grasping, manipulating, sealing, and dividing large tissue bundles.Do not bend instrument shaft.Do not attempt to seal over clips or staples as incomplete seals/damage to the cutting blade will occur.Contact between an active electrode and any metal objects may result in alternate site burns or incomplete seals.Do not overfill the jaws of the instrument with tissue, as this may reduce device performance.Keep the instrument jaws clean.Build-up of eschar may reduce the seal and/or cutting effectiveness.Wipe jaw surfaces and edges with a wet gauze pad as needed.The lever must be continually held with the activation button fully depressed until the seal cycle is complete.The lever does not latch into the activation position.A continuous tone sounds to indicate the activation of rf energy.When the activation cycle is complete, a two-pulsed seal-cycle-complete tone sounds and rf output ceases.Notice ¿ the surgeon may inspect the seal before cutting the vessel or tissue.After inspecting the seal, the surgeon should create a second seal.Adjacent to the first seal before cutting, as described below.A tone with multiple pulses indicates that the seal cycle was not completed.Refer to the troubleshooting section for possible causes and corrective actions.Do not cut tissue until you have verified that.There is an adequate seal.To seal adjacent tissue, overlap the edge of the existing seal.The second seal should be distal to the first seal to increase seal margin.Activating energy delivery with a footswitch when the activation button is not fully depressed may result in improper sealing and increase thermal spread to tissue outside the surgical site.Proper pressure is being applied to the tissue when the lever keeps the activation button fully depressed.Energy-based devices, such as esu pencils or ultrasonic scalpels that are associated with thermal spread should not be used to transect seals.Failure to maintain steady pressure on the lever while cutting can result in inadvertent reactivation of energy.Remove any embedded tissue from blade track and jaw hinge area.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, R
Manufacturer (Section D)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer (Section G)
STRYKER-ENDOSCOPY LAKELAND 5300
5300 region ct
lakeland FL 33815
Manufacturer Contact
andy dobos
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key16572147
Report Number0001056128-2023-00003
Device Sequence Number7423317
Product Code NUJ
UDI-Device Identifier07613327417845
UDI-Public07613327417845
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K180451
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberLF4418
Device Catalogue NumberLF4418RR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/20/2023
Initial Report FDA Received Date03/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-