| Model Number |
11330003 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
During investigation of the event, a failure of the xtc (tube cooling unit) was identified.The service engineer repaired the system by replacing the malfunctioning xtc component.The system was brought back to clinical operation.The exchanged component was returned for further hardware investigation.Initial analysis has determined that the affected part was not exchanged within the past three months considering the whole installed basis and no other complaints exist regarding this component.A supplemental report will be provided if the hardware investigation shows any systematic or design issues.
|
| |
|
Event or Problem Description
|
|
It was reported to siemens that a malfunction occurred while operating the neotoma alpha ct system.After a topo scan of a 3 year old male patient, when loading the next scan, the loading process was aborted with an error.The user restarted the ct system and repeated the topogram scan without success.The patient was then examined on another ct system.The patient was sedated during this procedure to perform the examination.No negative health consequences were reported for the child associated with the reported event.Therefore, this report has been submitted with an abundance of caution as siemens considers children as a vulnerable patient group.
|
| |
|