MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS BD¿ LUER SLIP TIP SYRINGE STERILE, SINGLE USE,; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS BD¿ LUER SLIP TIP SYRINGE STERILE, SINGLE USE,; PISTON SYRINGE Back to Search Results
Model Number 309659
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event is unknown; awareness date has been used for this.Initial reporter facility name: (b)(6).A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using bd¿ luer slip tip syringe sterile, single use, the plunger is separating from the stopper.There was no report of patient impact.The following information was provided by the initial reporter: the plunger is disconnecting from the stopper.
 
Additional Manufacturer Narrative
The following fields have been updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 28-mar-2023.H.6.Investigation summary: four sealed samples and one photo were provided to our quality team for investigation.The reported issue was not confirmed upon inspection of the samples.None of the samples showed any signs of the reported defect.Upon examination of the returned photo, it was observed that the stopper was disconnected from the plunger and was inside the syringe.Potential root cause for the stopper not seated on plunger defect is associated with the assembly process.These defects are occurring below an expected rate so no corrective actions will be made at this time.Furthermore, a device history record review was completed for provided lot number 2304034.A review showed no rejected inspections or quality issues during the production that could have contributed to the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10.
 
Event or Problem Description
It was reported while using bd¿ luer slip tip syringe sterile, single use, the plunger is separating from the stopper.There was no report of patient impact.The following information was provided by the initial reporter: the plunger is disconnecting from the stopper.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD¿ LUER SLIP TIP SYRINGE STERILE, SINGLE USE,
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16573826
Report Number1213809-2023-00221
Device Sequence Number5980183
Product Code FMF
UDI-Device Identifier30382903096597
UDI-Public(01)30382903096597
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K110771
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number309659
Device Catalogue Number309659
Device Lot Number2304034
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/01/2023
Supplement Date Received by Manufacturer04/13/2023
Initial Report FDA Received Date03/20/2023
Supplement Report FDA Received Date04/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-