Model Number 3000
Medical Device Problem Code Gradient Increase (1270)
Health Effect - Clinical Codes Chest Pain (1776); Dyspnea (1816); Heart Failure/Congestive Heart Failure (4446)
Date of Event 02/03/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional narratives: stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient related contributing factors.Structural valve deterioration (svd), a common reason for bioprosthesis explant or reoperation, encompasses multiple failure modes, including calcific and non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors, and is not usually an indication of a device malfunction.The device was not returned for evaluation, as device request is ongoing.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Event or Problem Description
It was learned via the implant patient registry that a 21mm bioprosthetic aortic valve, implanted for six (6) years and four (4) months, was explanted due to stenosis.The patient presented with hf, sob, and chest pain.The explanted device was replaced with a 21mm aortic valved conduit.Per medical records the preoperative diagnosis was prosthetic valve dysfunction and moderate non ischemic cardiomyopathy.Preoperatively the patient was diagnosed during work up with a stove pipe aorta.The patient underwent cardiac decortication, a modified bentall redo avr procedure using a 21 mm aortic valved conduit and cabg x 1.The patients post operative course was complicated by respiratory failure, aki and suspected hit.The patients continued to deteriorate and the patient was transitioned to comfort care.The patient expired on pod #15.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital and patient family members) and is not received in the form of a conventional "customer complaint." the information reported may or may not be related to the edwards device.
 
Additional Manufacturer Narrative
The most likely cause is patient factors, including hyperlipidemia (hld) and diabetes mellitus (dm).The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA AORTIC BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
one edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key16577063
Report Number2015691-2023-11603
Device Sequence Number4642784
Product Code DYE
UDI-Device Identifier00690103162240
UDI-Public(01)00690103162240(17)190911
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/11/2019
Device Model Number3000
Device Catalogue Number300021MM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2023
Supplement Date Received by Manufacturer04/18/2023
Initial Report FDA Received Date03/20/2023
Supplement Report FDA Received Date04/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/12/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient Age60 YR
Patient SexFemale
-
-