MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMPLATZ SUPER STIFF; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMPLATZ SUPER STIFF; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 39575
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a foreign material was found on the device.A 035/260 amplatz super stiff guidewire was selected for use.However, during preparation, a foreign matter was found in individual product bag.The procedure was completed with another of the same device.There were no patient complications reported.
 
Event or Problem Description
It was reported that a foreign material was found on the device.A 035/260 amplatz super stiff guidewire was selected for use.However, during preparation, a foreign matter was found in individual product bag.The procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
The device was inspected according to procedure.The device returned inside the pouch, and it was observed that the pouch information matches with the complaint information.The pouch was closed, and no open section was identified, however, a foreign matter was observed inside the package.There are none visual defects, contamination or damages observed on the pouch.The foreign matter was measured with the size estimation chart.According to the procedure, the foreign matter does not exceed the maximum size allowed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZ SUPER STIFF
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jeff wallner
4100 hamline ave n
arden hills, MN 55112
6515811560
MDR Report Key16577164
Report Number2124215-2023-09787
Device Sequence Number4697149
Product Code DQX
UDI-Device Identifier08714729115830
UDI-Public08714729115830
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K843016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/15/2025
Device Model Number39575
Device Catalogue Number39575
Device Lot Number0030728056
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/24/2023
Supplement Date Received by Manufacturer04/05/2023
Initial Report FDA Received Date03/20/2023
Supplement Report FDA Received Date05/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-