MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TI LOCK-SCR CANCLS 6.5X20MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TI LOCK-SCR CANCLS 6.5X20MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Osteopenia/ Osteoporosis (2651); Unequal Limb Length (4534)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034 -2023 -00580; 0001825034 - 2023 -00581; 0001825034 - 2023 -00582; 0001825034 - 2023 -00583; 0001825034- 2023-00584; 0001825034 - 2023 -00585; 0001825034 - 2023 - 00586.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported by the pmi group that a patient underwent right hip arthroplasty on an unknown date.Subsequently, the patient is being considered for a pmi product on an unknown day due to leg length discrepancy.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further information at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected} updated: d1; d4; d10; g3; g4; h2 d10: cat# xl-108525 lot# 832500 arcomxl 40mm +3 hw lnr sz 25 cat# 11-300916 lot# 937720 arcos 16x190mm spl tpr dist cat# 11-301324 lot# 836460 arcos con sz d std 70mm cat# 010001036 lot# 6726723 cocr fem hd 40mm type 1 -3mm cat# 00223200418 lot# 64629062 cable cerclage cable with crimp 1.8 mm dia.635 mm length it was reported that one of the seven screws fractured.It is unknown which screw has fractured.
 
Event or Problem Description
No further information at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: the acetabular implant is loose and displaced superiorly as noted.Bone quality is osteopenic.All of the screws appear loose and an inferior screw is fractured.The resulting superior displacement of the arthroplasty would result in a limb length discrepancy.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI LOCK-SCR CANCLS 6.5X20MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16582230
Report Number0001825034-2023-00587
Device Sequence Number20087231
Product Code LPH
UDI-Device Identifier00880304521209
UDI-Public(01)00880304521209(17)291217(10)263990
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K991807
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberCP161941
Device Lot Number263990
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2023
Supplement Date Received by Manufacturer04/03/2023
07/12/2023
Initial Report FDA Received Date03/21/2023
Supplement Report FDA Received Date04/19/2023
07/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/17/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Other;
Patient Age54 YR
Patient SexMale
-
-