| Model Number |
CA261TS |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
Sore Throat (2396)
|
| Date of Event |
03/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient has alleged to experiencing throat irritation.There was no report of serious patient harm or injury.There was no medical intervention required by the patient.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer received new and relevant information on 03/23/2026.A device was returned to a third party service center.During the evaluation of the device at the third party service center, the device was visually inspected and visible foam particles were observed.Additionally, the patient¿s complaint was confirmed.Section "product udi" g3 and h6 have been updated in this follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
The following correction has been updated in this report.Section "evaluation method code grid" in box h6 has been updated/corrected." the affected products with field complaints alleging pe-pur foam degradation have undergone the establishment of complaint codes specifically identifying foam degradation, for system one.A field correction action (2021-06-a) was initiated, and medical device recall letters were issued.A field action was initiated to replace pe-pur foam in affected products; degradation was added to the respective dfmeas and the risk management file ¿ (b)(4) v27 (systemone) was updated to include hazards associated with foam degradation based on complaints analyzed through various health hazard evaluations (hhes).Risk tags ¿ degradation01, toxic02 and particulate01 reflect the safety risk of design containing the pe-pur foam.These complaints were monitored and trended in accordance with the complaint trending procedures.If the complaints were determined to meet the criteria for escalation, they were escalated for risk assessment through the post market clinical escalation process.As of the date of submission for this report, there is no indication that complaints for the failure in question have led to unacceptable levels of severity or probabilities of harm, and therefore no further action will be pursued.".
|
| |
|
Search Alerts/Recalls
|