• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA STEM: MASTERLOC CEMENTLESS TI COATED LAT STEM SIZE 8; HIP STEM CEMENTLESS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA STEM: MASTERLOC CEMENTLESS TI COATED LAT STEM SIZE 8; HIP STEM CEMENTLESS Back to Search Results
Model Number 01.39.208
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Implant Pain (4561)
Date of Event 02/21/2023
Type of Reportable Event Serious Injury
Event or Problem Description
At about 5 years 1 month after the primary, the patient came in reporting pain due to an unstable hip.There was anteversion of the stem.The surgeon revised the medacta stem and head with competitor components and revised the medacta liner with a medacta liner.The surgery was completed successfully.
 
Additional Manufacturer Narrative
Batch review performed on 21 march 2023.Lot 168824: (b)(4) items manufactured and released on 05-may-2017.Expiration date: 2022-04-12.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM: MASTERLOC CEMENTLESS TI COATED LAT STEM SIZE 8
Common Device Name
HIP STEM CEMENTLESS
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco
strada regina
MDR Report Key16583376
Report Number3005180920-2023-00181
Device Sequence Number20064918
Product Code LZO
UDI-Device Identifier07630030885686
UDI-Public07630030885686
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K151531
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/12/2022
Device Model Number01.39.208
Device Catalogue Number01.39.208
Device Lot Number168824
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2023
Initial Report FDA Received Date03/21/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/05/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-