• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3543207MC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 03/06/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture, breast prosthesis rupture.Explantation month, day, and year: (b)(6) 2023.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a caucasian female patient who underwent a primary breast augmentation procedure with mentor memorygel breast implant 320cc gel breast prostheses developed baker grade iv capsular contracture in both of her breasts post implantation.Additionally, it was reported that both of the patient¿s breast prostheses ruptured post implantation.Bilateral rupture was diagnosed via mri.As a result, the patient is scheduled to undergo explantation on (b)(6) 2023.This medwatch report is for the patient¿s left breast prosthesis.
 
Additional Manufacturer Narrative
On 24-may-2023, mentor received additional information about the event.The patient age at the time of event is 66 years old.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On 26-jun-2023, mentor received additional information about the event.The patient¿s explanted breast prostheses were replaced with mentor memorygel xtra breast implant 295cc gel breast prostheses.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On 20-jun-2023, the mentor failure analysis lab received the device for evaluation.On 22-jun-2023, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced a rupture in the breast implant and developed capsular contracture.The product was returned to mentor for evaluation.Mentor conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample revealed that the breast implant had an area of siltex cracking on the posterior view.In addition, a tear was noted within the siltex cracking measuring approximately 0.7 cm.A microscopic examination was performed and instrument damage was not found on the area of the tear.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.Mentor concluded that the capsular contracture in the patient¿s breast was the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsules, and effusion to pathology for examination.It should be noted that as part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key16584548
Report Number1645337-2023-03251
Device Sequence Number1583783
Product Code FTR
UDI-Device Identifier00081317004400
UDI-Public00081317004400
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2018
Device Model Number3543207MC
Device Catalogue Number3543207MC
Device Lot Number6680633
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/07/2023
Supplement Date Received by Manufacturer05/24/2023
06/20/2023
06/26/2023
Initial Report FDA Received Date03/21/2023
Supplement Report FDA Received Date06/05/2023
06/28/2023
07/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/20/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-