| Model Number |
00884450646533 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during evaluation of an artery in the patient's upper extremity for dialysis treatment, the tip of the catheter detached within the brachial artery.A simple benston wire was used for deliverability.A stiff guidwire was not used for this procedure.No tortuous, diseased, or calcified arteries were noted during the procedure, the detached catheter tip was easily retrieved with a snare.No additional patient consequences to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.The complaint was confirmed, and the root cause is attributed to clinical use and forces imposed during use.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|