MAUDE Adverse Event Report: CARDINAL HEALTH 12 FR X 55 ENTRIFLEX W/STYLET; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH 12 FR X 55 ENTRIFLEX W/STYLET; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884721255
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Dyspnea (1816); Foreign Body In Patient (2687)
Date of Event 03/14/2023
Type of Reportable Event Serious Injury
Event or Problem Description
The customer reported that the gravity head of the nasal feeding tube broke in the patient's esophagus in the department of general surgery of henan cancer hospital, which caused a great impact on the patient and department.The patient had shortness of breath after the tube was placed, and was immediately sent to scanning imaging.Later, the gravity head was found to have fallen in the airway, so it was taken out in time and observed.Per additional information provided on (b)(6) 2023, the weighted tip detached during the placement of the ng tube.On the afternoon of (b)(6) 2023, a ct scan of the patient's abdomen and chest was performed, looking for the tip of the weight until found, it was located in the patient's trachea.There were uncomfortable symptoms of respiratory distress during the period.The detached heavy tip was removed by using a tracheoscope clamp.A new ng tube was placed after the removal of the tip.The patient is in good condition currently.
 
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Additional Manufacturer Narrative
Section h6 updated to include impact code 4627: device explantation.
 
Additional Manufacturer Narrative
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The device was manufactured on july 14, 2020.Two photos and one physical sample were received for investigation.The photos and sample were evaluated, and the reported condition was confirmed.A corrective and preventative action has been initiated to further investigate and address the reported condition.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12 FR X 55 ENTRIFLEX W/STYLET
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
MX  02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key16586508
Report Number9612030-2023-03648
Device Sequence Number14683848
Product Code KNT
UDI-Device Identifier10884521518926
UDI-Public10884521518926
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Expiration Date06/30/2025
Device Model Number8884721255
Device Catalogue Number8884721255
Device Lot Number2019611464
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/15/2023
Supplement Date Received by Manufacturer03/15/2023
03/15/2023
Initial Report FDA Received Date03/21/2023
Supplement Report FDA Received Date03/29/2023
05/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-