| Model Number |
68LP-400 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device as been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: observed, one opening on posterior side assessed as fold flaw opening.Additional observations: ¿ wear abrasion is observed.¿ creases are observed.¿ yellow particles on device inner surface are observed.No further actions are required since no issue in the manufacturing of the device is observed.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|