| Model Number |
20030 REV 09, 20040 REV 08 |
| Medical Device Problem Code |
Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Uvision360 (hereafter uvision) received the returned product and found that the customer sent back a non-luminelle power cable and hdmi cable.Uvision performed a functional investigation and determined that when utilizing the luminelle power cable and hdmi cable, the device functioned as intended.When functional testing was performed on the power cable and hdmi cable the customer sent back, the device did not function as intended.Uvision did not consider this a medical device reportable event and therefore did not report this complaint when initially received due to the following: - after investigation and inspections of the returned product, it was found that the non-luminelle power cable being used with the device did not provide sufficient power to the device.- no patient harm was observed as a result of the malfunction.- based on the medical advisor's qualified opinion, this malfunction was unlikely to cause serious harm or death to a patient if this malfunction were to recur.- the malfunction was evaluated as a possible hazardous situation and failure mode of the device through the product hazard analysis and dfmea.All risks associated with this malfunction were deemed acceptable.- the maude database was reviewed for the product code hih and it was found that no events were reported for similar devices, as defined by fda guidance "medical device reporting for manufacturers guidance for industry and food and drug administration staff 2016".Following an fda routine inspection performed in februrary 2023, the inspector issued an fda 483 identifying this complaint as a reportable malfunction.
|
| |
|
Event or Problem Description
|
|
The customer reported during a phone call that they were unable to produce an image in their monitor after assembling the system.The customer indicated that the system failure was caught prior to be being used with the patient.
|
| |
|
Search Alerts/Recalls
|