| Model Number |
MAJ-1511 |
| Medical Device Problem Code |
Device Reprocessing Problem (1091)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported to olympus that during reprocessing, the connecting tube was popping off.The connecting tube could not be connected to the channel connector in the reprocessing basin.There were no reports of patient harm associated with this event.Subsequently, the customer reported that the device is not being returned because the problem was resolved.The issue was due to user error.This complaint is related to patient identifier (b)(6) (lf-gp/serial number: unknown).
|
| |
|
Additional Manufacturer Narrative
|
|
The subject device referenced in this report is currently working correctly and will not be returned for evaluation; therefore, the device evaluation could not be completed.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide additional information.
|
| |
|
Event or Problem Description
|
|
This complaint is related to patient identifier (b)(6) (oer-pro/endoscope reprocessor/serial number: unknown).
|
| |
|
Additional Manufacturer Narrative
|
|
Updated: d8, h6.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Section g2 was corrected to additionally indicate that the initial reporter is a healthcare professional.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the connecting tube was due to user error.It is possible that the tube was not pushed enough until it clicked during connection.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿over time, the connecting tube will deteriorate because it is subject to wear with each use.If any of the following are found with the connecting tube, do not use it and replace it with a new one.Connecting a defective tube could prevent the effectiveness of the cleaning and high-level disinfection process.Move the lock levers of the connecting tube to make sure that they function properly and are not broken.Connect each connector firmly by pushing until the connector clicks into place.After connection, pull the connector gently to confirm that it cannot be disconnected easily.¿ olympus will continue to monitor field performance for this device.
|
| |
|