The device model kd-v411m-0330 was returned for evaluation.The lot number was 27k with supplementary information number of ¿01¿.(m-bc manufacture date: july 1, 2022).Device evaluation, the following were noted: the cutting wire was broken.The investigation was performed to confirm the condition of the breaking portion.It was scorched and melted.In investigation performed in the past, a coated portion damage was duplicated by heat generation due to an electrical discharge.Shape of the coated portion of the subject device was similar to the shape confirmed in the past investigation result.The outer diameter of the cutting wire was measured.The result indicated no abnormalities.The length of the coated portion of the cutting wire, and the cutting wire itself presented no abnormalities.There were no missing parts in the subject device.Other abnormalities that could lead to the breakage of the cutting wire were not confirmed.The device history records (dhrs) for this product have been reviewed.No abnormalities were detected in the device history record with the lot number for the following inspection items, which related to the reported phenomenon.Process inspection sheet.Quality inspection sheet.Nonconforming product report.Ifu (instruction for use): this instruction manual contains the following information.(drawing no.Gk6224 revision no.9).Since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.When inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.Do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.If you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.Based on the results of confirmation of the device and the investigation results in the past, a likely mechanism causing the broken cutting wire might be the following.The device was not protruded enough from the endoscope until the rear end of the cutting wire was in the field of view.Due to the situation of ¿1¿ description, the cutting wire and the endoscope were being close to each other.The output was activated in state of ¿2¿ description.This might have led to an electrical discharge between the cutting wire and the distal end of the endoscope.An electrical discharge possibly occurred, and the cutting wire became hot instantly.That might have caused the cutting wire to break.It can be inferred that heat generated by an electrical discharge caused damage of the coated portion.Olympus will continue to monitor complaints for this device.
|
|
As reported, during a therapeutic endoscopic sphincterotomy (est) procedure, knife wire breakage occurred at the first energization.The broken wire did not fall out of the body.The device was replaced with similar device and the intended procedure was completed with no harm or injury to the patient.No user injury reported.No harm was reported.No patient harm, no user injury reported.
|