• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG CONSTRAINED INSERT 28F; PROSTHESIS, HIP, CONSTRAINED, CEMENTED OR UNCEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0 DEG CONSTRAINED INSERT 28F; PROSTHESIS, HIP, CONSTRAINED, CEMENTED OR UNCEMENTED, METAL/POLYMER Back to Search Results
Model Number 690-00-28F
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned to the manufacturer.
 
Event or Problem Description
The following was reported: patient had second hip revision procedure in two months.Her third hip revision procedure in total.On this latest occasion, her hip prosthesis had dislocated similar to the previous event.Specifically, she had a trident constrained insert in-situ and the bipolar head had dislocated from the insert that it had been pre-assembled in.The insert was still in the acetabular shell but the bipolar head was completely disengaged from it.The bipolar head still had the femoral head within it which itself was still intact on the cone body implant trunnion.
 
Event or Problem Description
The following was reported: patient had second hip revision procedure in two months.Her third hip revision procedure in total.On this latest occasion, her hip prosthesis had dislocated similar to the previous event.Specifically, she had a trident constrained insert in-situ and the bipolar head had dislocated from the insert that it had been pre-assembled in.The insert was still in the acetabular shell but the bipolar head was completely disengaged from it.The bipolar head still had the femoral head within it which itself was still intact on the cone body implant trunnion.Update: as per intake form, the surgeon explanted the trident constrained liner, ceramic head, trident cluster shell, screws and the cone body.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation involving a trident liner ¿was reported.¿¿the event was confirmed as the bipolar head containing the ceramic head was returned independent to the constrained liner.Method & results: -device evaluation and results: visual inspection of the returned device indicated damage consistent with explantation.The ring disassociated from the insert along with a screw still assembled.-clinician review: no medical records were received for review with a clinical consultant.-device history review: review of the product history records indicate devices were manufactured and accepted into final stock with no reported relevant discrepancies.-complaint history review: there has been no other event for the lot referenced.Conclusion: the event was confirmed as the bipolar head containing the ceramic head was returned independent to the constrained liner.Further information such as pre- and post-operative x-rays, operative reports as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0 DEG CONSTRAINED INSERT 28F
Common Device Name
PROSTHESIS, HIP, CONSTRAINED, CEMENTED OR UNCEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16590148
Report Number0002249697-2023-00287
Device Sequence Number4937751
Product Code KWZ
UDI-Device Identifier07613327025552
UDI-Public07613327025552
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K061654
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number690-00-28F
Device Catalogue Number690-00-28F
Device Lot Number6T3AWR
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/24/2023
Supplement Date Received by Manufacturer06/06/2023
Initial Report FDA Received Date03/22/2023
Supplement Report FDA Received Date06/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age81 YR
Patient SexFemale
-
-