Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Investigation summary: the product was returned for evaluation.Visual inspection evaluation was conducted on the returned device.Visual analysis of the returned sample determined that one detached needle and some suture pieces on the degradation process that pertains to the product code d10058 were received for evaluation.The product is a control release; each packet should contain eight strands.Upon visual assessment of the suture, it was observed that the strand have begun with a degradation process caused by exposure to the environment.This condition contributes to the needle separating from the suture.The foil related to (b)(4) was visually inspected, and excessive wrinkles and holes in the cavity were observed due to handling.Per the sample condition, the functional test cannot be performed.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2023 and suture was used.After opening the package before use, it was found that 2 needles had been detached from 2 sutures.Visual inspection evaluation was conducted on the returned device.Visual analysis of the returned sample determined that one detached needle and some suture pieces on the degradation process that pertains to the product code d10058 were received for evaluation.The product is a control release; each packet should contain eight strands.Upon visual assessment of the suture, it was observed that the strand have begun with a degradation process caused by exposure to the environment.This condition contributes to the needle separating from the suture.No adverse patient consequences were reported.Additional information was requested.
|